|
Fluorescence
Spectroscopy
Experimental
Biochemistry I Fluorospectroscopy Instructions
1. Turn on the POWER
switch (not Main Power yet).
2. After waiting 10 seconds,
press and hold the Xe Lamp START Button (located
next to the Main Power switch) for 3 or 4 seconds (do not hold it
in for more than 5 seconds). This should
turn on the Xe lamp. If the lamp turns on,
the Xe orange light will be dimly lit. The Xe lamp lamp cannot be
started while the Main switch is turned
on.
3. Turn on the MAIN POWER
switch. After a few seconds, the following screen
(in orange font color) should be displayed:
Initialization
ROM Check OK
RAM Check OK
.
.
.
4. After the instrument
goes through its initialization sequence, you are ready to
set the parameters you will need for the experiment.
5. Using the photometry
mode, make sure to set the DATA MODE to CONC. This
will allow you to measure the standard and analyte samples, and
concentration calculations are carried out.
6. Set up the sample
ID number (1 to 9999), and set the replicates to "3."
7. Under TEST SETUP,
make sure to set the HI LIMIT and LO LIMIT, as well as
the UNIT LABEL (mg/mL, ng/ML etc.) values.
8. Next, input the number
of standards (2 to 10) under the NUM STDS option.
9. Under CURVE MODE,
choose NEW, so a new curve will be developed as you
perform the experiment.
10. For CURVE DATA, also
choose NEW.
11. For the TEST NAME
option, enter a test name (refer to Table 1).
12. For INSTR SETUP,
set RESPONSE to 0.5 sec, BANDPASS to 10 nm, PM
VOLTAGE to 400, and TEXT PRINT to on.
13. Make sure to save
the parameters you have selected using the SAVE PARAMS
option.
14. If a wavelength to
be measured is not known, do a pre-scan (see Table 2)
to see what wavelength to use. If the wavelength is known, continue
to step 15.
15. After setting all
of the parameters, and knowing the wavelength to use, press
the FORWARD key, and the MEASUREMENT screen will appear.
16. Next, set the standard
sample 1, and press START (EX or EM) key.
17. Set standard sample
2, and then press the START key again.
18. After completing
the standard measurements, the results are reported on the
screen, and a calibration curve is developed.
19. Set the analyte sample,
and press the START key.
20. After measuring the
analyte sample, the measurement value and calculated
concentration will be displayed on the screen.
|