|
Atomic
Absorption Spectroscopy
|
|
|
Perkin-Elmer
AAnalyst 100
|
Operating
Instructions for Perkin-Elmer AAnalyst 100:
Flame Atomic Absorption Photometry
1. Turn on atomic absorption
unit (switch is on right hand side of instrument).
2. Turn on HGA 850 Graphite
Furnace (the switch is on the left hand side of instrument).
3. Turn on acetylene
tank by twisting the main valve ½ to one full turn (left
regulator should read around 15).
4. Turn on computer.
-When the window Enter Network Password
appears - click on Cancel .
-Double click on AA WinLab Analyst
icon.
-Checking Connections window will
appear then close if all is OK.
-When the window AA WinLab appears
check that the Technique
is
Flame .
-Click on button displaying green
Erlenmeyer flask (Use basic system procedure
for performing analyses with manual sampling).
-When the window Open Method appears:
Fill in Method name ("Ca in water"
for example).
-Click on New Elem. (right side of
window).
-When the window Create New Method
appears, click on down arrow 6 and
scroll through to find element that is going to be analyzed then
click on it. The element should appear
in the box next to the heading Element to be determined.
Click on OK . The window Manual Analysis will appear.
A. To align the lamp:
1. Check to see that the proper hollow
cathode lamp is installed (right side of
instrument under black door). If the wrong lamp is installed, carefully
remove lamp
and place it back into its box, then carefully place the correct
lamp into the instrument. (Make certain that the lamp is properly
seated into
its holder.)
2.
Click on the button in the AA WinLab window labeled Lamps.
3.
When the window Align Lamps appears check that Elements box and
the Lamp Type
boxes contain the correct information then click on the gray
button under Set Up.
4.
Close window Align Lamps.
B. To turn on the flame:
1. Make sure that the aspirating tube
is in distilled water.
2.
Click on the button in the AA WinLab window labeled Flame .
3.
When the window Flame Control appears make sure the Safety Interlock
box is green with a check.
4.
Click the switch that appears in the Flame Control window form Off
to On.
5.
Adjust the flame manually using fuel and oxidant valves on the front
of theinstrument
(the gauges should read about 3 for fuel and 8 for oxidant).
C. To adjust and optimize
burner:
1. Prepare a blank solution (usually
distilled water) and a sample containing
an ion to be analyzed (for Ca tap water is fine).
2.
Click on Tools a menu will appear, select and click on Continuous
Graphics.
3.
Aspirate blank when signal is steady click on Auto-zero.
4.
Aspirate the sample containing the analyte.
5.
Then use the three adjusting knobs on the burner to maximize the
signal, be
sure that when you adjust the height of the burner that you don't
block the optical path with the burner. To determine if the path
is blocked,
check to see if the signal goes back to zero when you place the
aspirating tube into distilled water.
D. To create a standard
curve:
1. Prepare standards of know concentration.
2.
Click on the button in the AA WinLab window labeled MethEd.
3.
Type in Method Desc. (Calcium in tap water) Check that the other
instrument
settings are OK.
4.
Click on the tab (lower left corner) Calib.. Also, make sure you
are on the
Equation, Units & Replicates tab/sheet. Enter the appropriate
information:
Calibration
Equation
Equation
Zero Intercept: Linear
Max.
Decimal Places: 3 Max. Significant Figures: 4
Units
Calibration
. choose appropriate units
Sample
.
choose appropriate units
Replicates
Number
Fixed 3
5.
Click on the tab Standard Concentrations (upper right side). Enter
information
for each of your standards - standard ID and concentration. Note:
Do NOT enter zero as a standard concentration. The blank is designed
to be used in the place of the zero concentration standard.
Go to
Manual Analysis - Aspirate distilled water or appropriate blank
and click
on Analyze Blank Click on Results icon
6. To analyze standards, use the pull-down
menu to choose the correct standard
and click on Analyze Standard.
-Click on
calibration button to show calibration curve - After each standard is
run, rinse sample capillary with distilled water.
-Go to Sample
Editor
Enter data
about samples
-Go to Manual
Analysis
-Click on
analyze sample
7.
Print sample information
-Select the
windows you would like to print - like your calibration curve and
then choose print - graphics
E. Shutting down the
instrument:
1. When analysis is complete
-Click on
flame icon and shut off flame
-Go to file
- exit - OK to lose method
2.
Shut off computer
-Next, turn
of the instrument using the switch
-Finally,
turn off the Acetylene tank
Operating
Instructions for Perkin-Elmer AAnalyst 100:
Graphite Furnace (HGA 850) Usage
Getting Started:
1. Turn on atomic absorption
unit (switch is on right hand side of instrument).
2. Turn on HGA 850 Graphite
Furnace (the switch is on the left hand side of
instrument).
3. Verify that the fume
vent is operational and turned on.
4. Turn on argon gas
(and special gas) to the specified pressures (see pressure
guides Table 1 and Table 2).
5. Turn on the cooling
system by pressing the power switch at the rear panel
of the device. The cooling system is the device that looks like
a CPU, located below the lab bench.
-Normally, it should
be full to the maximum amount of coolant. The coolant
is a special mix of a 1:10 solution of glycerol to water (1 part
glycerol and
9 parts water).
6. Turn on the computer.
-When the window Enter Network Password
appears, click on Cancel .
-Double click on the AA WinLab Analyst
icon.
-The Checking Connections window will
appear, and then close it if everything
is OK.
- In the File menu, click on Technique. In the Select Technique
dialog, click
on Furnace and then on OK.
-When
the window Open Method appears:
-Fill
in Method name ("Ca in water" for example).
-Click
on New Elem. (right side of window).
-When
the window Create New Method appears, click on down arrow 6
and scroll through to find element that is going to be analyzed
then click
on it. The element should appear in the box next to the heading
Element
to be determined. Click on OK.
-The
window Manual Analysis will appear.
7. To align the lamp:
1) Check to see that the proper hollow
cathode lamp is installed (right side
of instrument under black door). If the wrong lamp is installed,
carefully
remove lamp and place it back into its box, then carefully place
the correct lamp into the instrument. (Make certain that the lamp
is properly seated into its holder.)
2)
Click on the button in the AA WinLab window labeled Lamps .
3)
When the window Align Lamps appears check that Elements box and
the Lamp Type boxes contain the correct information then click on
the gray button under Set Up.
4)
Close window Align Lamps.
8. Optimize the position
of the furnace:
1) Make sure that the lamp for the
element of study is already installed and aligned
properly.
2)
Switch off the deuterium lamp of the background corrector if it
is on.
3)
Move the furnace out of the atomizer compartment.
4)
Verify that the graphite tube intended for use is installed in the
furnace, after
removing the window assemblies.
5)
Autozero the spectrometer:
- In the Tools
menu, click on Continuous Graphics.
- Click on
Autozero.
6)
Move the furnace back into the atomizer compartment. The absorbance
reading should
be less than 0.002. If it isn't, then align the furnace according
to the steps below:
-
Move the locking lever to the left in order release lateral (left
to right)
movement of the furnace.
- Rotate the vertical
(up and down) and horizontal (forward and backward)
alignment controls alternately and in very small increments
to find the position giving minimum (less than 0.002)
absorbance reading on the display.
- Holding the furnace
assembly at the base, carefully move the furnace laterally
to the left and to the right to find the position that gives minimum
absorbance reading (less than 0.002) on the display.
- Repeat the horizontal,
vertical, and lateral adjustments until you have achieved
the minimum (less than 0.002) absorbance reading.
7)
After aligning the furnace:
- In the Tools
menu, click on Continuous Graphics.
- Click on
Autozero.
9. Reset the window assemblies.
The absorbance, after replacing the windows,
should be between 0.06 to 0.09 measured at the Cu line at 324.8
nm (for example).
10. Conditioning the
graphite furnace:
1) Make sure that the graphite tube
is properly installed and already aligned.
2)
Click on Furnace on the Toolbar to open the Furnace Control Window.
3)
Within the Furnace Control window, click on Condition Tube. The
system uses
a special furnace program for heating the tube to successively
higher temperatures.
|